SciELO - Scientific Electronic Library Online

 
vol.56 número3CHARACTERIZATION OF HYDRATED TITANIUM OXIDE WITH SHEET-LIKE AND TUBE-LIKE STRUCTURES PREPARED BY A HYDROTHERMAL PROCESSINGDETERMINATION OF 10 - HYDROXY - 2 - DECENOIC ACID IN ROYAL JELLY BY CAPILLARY ELECTROPHORESIS índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Journal of the Chilean Chemical Society

versión On-line ISSN 0717-9707

J. Chil. Chem. Soc. vol.56 no.3 Concepción  2011

http://dx.doi.org/10.4067/S0717-97072011000300003 

J. Chil. Chem. Soc., 56, N° 3 (2011), págs.: 734-737.

 

SIMULTANEOUS DETERMINATION OF ENALAPRIL AND STATIN'S IN PHARMACEUTICAL FORMULATIONS BY RP-HPLC

NAJMA SULTANA1, M. SAEED ARAYNE1, SAFILA NAVEED2

1 Dow University of Health Sciences, Karachi, Pakistan.
2 Faculty of Pharmacy, Hamdard University, Karachi- Pakistan.
e-mail: safila117@gmail.com


ABSTRACT

Simple, specific, economical and precise high performance liquid chromatographic method for the simultaneous determination of enalapril in presence of statins (rosuvastatin, atorvastatin and simvastatin) in API (active pharmaceutical ingredient) and formulation has been developed and validated. Chromatography was carried out at 25"C on a prepacked Purospher Star, C18 (5 mm, 250 x 4.6 mm) column with the isocratic mobile phase of acetonitrile: water (60:40 v/v) adjusting pH to 2.8. The UV detection was carried at 230 nm. The results obtained showed good agreement with the declared contents. Enalapril and statins separated in less than 10 mins with good resolution and minimal tailing and without interference of excepients. The method was linear in the range of 2.5-100 μgmL-1 for enalapril concentration with a correlation co-efficient 0.9995 and in the range 0.625-25 μgmL-1 for statins concentrations having correlation coefficient 0.9990 (inter and intraday CV<2 .0%). The recovery was 99-102%. The method was validated according to ICH guidelines and the acceptance criteria for accuracy, precision, linearity, specificity and system suitability were met in all cases. The proposed method can be used for quantitative determination of enalapril and statins alone or in combination from API and formulations.

Keywords: Enalapril; statins; rosuvastatin; atorvastatin; simvastatin and RP-HPLC.


 

INTRODUCTION

In literature, it has been shown that antihypertensive drugs and statins are effective in preventing cardiovascular disease1. Chemically enalapril is designated as (S)-1-[N-[1-(ethoxycarbonyl)-3-phenyl propyl]-L-alanyl]-L-proline,(Z)-2-butenedioate salt is a drug used in the treatment of hypertension and chronic congestive heart failure.2

Control of hypercholesterolemia is of prime importance for the primary and secondary prevention of coronary artery disease (CAD).3,4 3-hydroxy-3-methylglutarylcoenzymeA(HMG-Co A) reductase inhibitors (statins) are often prescribed in association with antihypertensiveagents5. Clinical trials have unequivocally demonstrated that treatment of dyslipidemias reduce cardiovascular (CV) evets6 . Various HPLC methods have been reported for estimation of enalapril maleate. Tajerzadeh et.al., 7 developed a method of enalapril maleate with UV detection having LOD of 0.125 and LOQ of 0.5 μgmL-1. Kyung et.al and Pisarev et.al, 9 also developed a method for enalapril using mass spectrometrometry. Several methods are also reported for the determination of statins 1012. Our research group has worked on the simultaneous determination of a number of commonly co-adminstered drugs as olmesartan medoxamil and irbesartan and hydrochlorothiazide13, atenolol, rosuvastatin, spironolactone, glibenclamide and naproxen sodium14, enalapril maleate and H2-receptor antagonists15, diltiazem and statins16, lisinopril and statins17, prazosine and statins18, ceftriaoxone sodium and statins19and antidiabetic drugs with statins20.

These methods can be applied for the quantitation of drugs as well as for clinical purposes. Other antihypertensive drugs such as amlodipine and statins are available in combined formulations. Simultaneous determinations of these combinations are also present in literature survey, but not a single method for the determination of enalapril and statins is available in literature although both of these drugs are given together. On this basis, it becomes apparent to develop and validate a simultaneous method for the determination of these drugs in bulk material and dosage formulation using RP-HPLC in a relatively short time with high linearity. Furthermore, this validated method was used to study the possible in vitro interactions of enalapril with statins (atorvastatin, rosuvastatin and simvastatin) at different simulating body environments, results of which are reported elsewhere.

EXPERIMENTAL

Instrumentation and chromatographic conditions

Shimadzu HPLC system equipped with LC-10 AT VP pump, DGU-14 AM on-line degasser, from Shimadzu Corporation (Chromatographic and Spectrophotometric Division, Kyoto, Japan) consisted of a ternary gradient system, column and UV detector. Isocratic mobile phase consisted of acetonitrile: water (60:40 v/v) filtered and degassed through membrane filter of 0.45 micron porosity. Purospher Star, C18 (5 mm, 250 x 4.6 mm) column was used as stationary phase. The flow rate was l.8 ml min-1 and the detector was set at 230 nm. All analyses were made at 25"C and the volume of solution injected was 20 μL .Chromatograms were recorded and integrated on PC installed with Shimadzu CLASS-GC software (Version 5.03).

Reference substances, reagents and chemicals

Reference standards of enalapril from BMS and statins (rosuvastatin, atorvastatin and simvastatin) were obtained from Pharm Evo (Pvt.) Ltd., Atco Pharma (Pvt.) Ltd., and Geofman Pharma (Pvt.) Ltd., respectively. Enalapril (renitec 10) was a gift from BMS Pvt Ltd. Statins used were rosuvastatin (X-plended 20mg), atorvastatin (Atopitar 10mg) and simvastatin (Atcol 10mg) which were purchased from the local pharmacy. Acetonitrile (HPLC grade) was obtained from Merck, Germany Deionizer; Stedec CSW-300 was used for deionization of water. All the chemicals and reagents were of analytical or reagent grade.

Solution preparation

Enalapril and statins standard stock solutions

Stock standard solutions of 100 μgmL-1 of enalapril, and 25 μgmL-1 simvastatin, atorvastatin, and rosuvastatin were prepared by dissolving 10 mg of enalapril and 2.5 mg of statins in 100 mL volumetric flask, 10 milliliters of diluent was added initially and sonicated for a few minutes to solubilize these drugs, then 30 ml of mobile phase was added, sonicated to dissolve drugs. The solution was diluted to volume with the mobile phase and mixed.

Standard solution

Solutions were also prepared by diluting these to obtain concentration between 2.5-100 μgmL-1 for enalpril and 0.25-25 μgmL-1 for statins. These solutions were stored at 20°C, they were prepared once and analyzed daily for inter-day and inter-operator variations of the method and analyzed each time before drug analysis in biological samples. 20 μL of these solutions were injected into LC system and chromatographed.

Determination from formulations

For testing the suitability of the proposed method for the estimation of the drugs in dosage forms, 20 tablets of each drug were powdered and equivalent to 2.5 g of statins (simvastatin, atorvastatin, rosuvastatin and pravastatin) and 10 mg of enalapril were transferred individually to 100 mL volumetric flask, dissolved and diluted with mobile phase. The resulting solutions were filtered through Whatman filter paper no. 41 and diluted to the desired concentration and analyzed for the drug content.

RESULTS AND DISCUSSION

Chromatographic condition

Initially, a Purospher Star, C18 (5 mm, 250 x 4.6 mm) column in isocratic mode, with mobile phase acetonitrile and water in proportion of 60:40 (v/v) at a flow rate of 1.8 ml/min at a detection wavelength of 230 nm was used. To optimize the operating conditions for isocratic RP-LC detection of all analytes, a number of parameters such as the mobile phase composition, pH and the flow rate were varied. Various ratios (80:20, 70:30, 60:40 v/v) of methanol: water was tested as starting solvent for system suitability study then acetonitrile and water having the above ratios was tried. The variation in the mobile phase leads to considerable changes in the chromatographic parameters, like peak symmetry, capacity factor and retention time. The pH effect showed that optimized conditions are reached when the pH value is 2.8, producing well resolved and sharp peaks for all drugs assayed. Henceforth, in the present method pH was adjusted to 2.8 using wavelength 230 nm (isobestic point). However, the peak shape and resolution were found to be good when the mobile phase comprising of the ACN: water having pH adjusted to 2.8 with phosphoric acid was used in the ratio of (60:40 v/v) at a flow rate of l.8 mLmin1 (filtered through a 0.45 micron filter). For simultaneous determination of enalapril with simvastatin, rosuvastatin, and atorvastatin individual drug solutions were injected into the column at the concentration of 100 μgmL-1 and both elution pattern and resolution parameters were studied. System suitability data is given in table 1. The retention time for enalapril was found to be 2.4 minute, rosuvastatin 3.3 minute, atorvastatin 4.0 minute and for simvastatin was 8.0 minute meeting the resolution criteria specified in USP 2008. A typical chromatogram of test solution is shown in figure 1.


Figure.1: A representative chromatogram of (a)Enalapril (b) rosuvastatin (c)atorvastatin and(d) simvastatin in raw material.

Method validation

The newly developed method has been validated and holds well for the determination of drug in raw materials, dosage formulations and serum. For validation of analytical methods, the guidelines of the International Conference on the Harmonization of Technical Requirements for the Registration of Pharmaceuticals for Human Use have recommended the accomplishment of system suitability, selectivity, specificity, linearity, accuracy test, precision, sensitivity, limit of detection and quantification of the method.

System suitability testing

Typical system suitability results are summarized in table 1, all the values for the system suitability parameters are within limits. The method was validated according to the ICH guidelines. 2,11


Selectivity and Specificity

The selectivity and specificity of the method was established through the study of resolution factor of the peak of enalapril from that of statins (table 1). The method demonstrated good resolutions and was found to be free of interference from the excipients (Fig. 2) used in formulation products and thus, the method is specific for enalapril and statins.


Linearity

Enalapril and statins showed linear calibration curves in the range of 2.5, 5, 10, 25, 50 and100 μgmL- for enalapril and 0.625, 1.25, 2.5, 6.25, 12.5 and 25 μgmL for statins respectively (r2 > 0.999). Table 2 shows the regression statistics of enalapril and statins.


Accuracy

The accuracy of the method was evaluated from the recovery results of spiked placebo samples. Appropriate portions of stock solution of enalapril and statins were spiked into blank placebo matrix to produce concentrations of 80, 100 and 120% of the theoretical concentration. Mean recovery of samples was 100.1% for enalapril and 99.40% for statins data given in table 3.


Precision

Precision was determined by six replicate determinations of standard solution and the relative standard deviations were <2% for enalapri and statins. Method precision or intra-assay precision was performed by preparing six different samples involving different weightings. Each solution was injected in triplicate under the same conditions and the mean values of peak area responses for each solution were taken. Intermediate precision was performed by analyzing the samples on two different days (table 4) employing different instruments.


Robustness

Robustness of the proposed method was estimated by changing:(i) mobile phase composition from methanol:water to acetonitrile:water (60:40 v/v); (ii)the pH (iii) the flow rate from 1.0 ml to 1.8ml min1. System suitability parameters in table 1 were found to be within acceptable limits.

CONCLUSIONS

In short, our method is specific, sensitive, rapid and easy to perform for simultaneous determination of enalapril and statins (rosuvastatin, atorvastatin and simvastatin). The limit of quantification, small sample volume and short chromatographic time of this method makes it advantageous for adaptation to routine assay requirements and enables simultaneous determination of enalapril and statins because of good separation of the chromatographic peaks. The obtained results are in good agreement with the declared contents of dosage formulations. Results are accurate and precise and are confirmed by the statistical parameters. Reliability, rapidness, simplicity, sensitivity, economical nature, good recovery and precision of this method give it advantage over the other reported methods.

REFERENCES

1. H. M. Colhoun, D.J. Betteridge, P.N. Durrington . The Lancet, 364, 685-96 (2004).         [ Links ]

2. G G Alfred, S G Louis, W R Theodore and M Ferid Goodman and Oilman's, McGraw Hill, 904, 1996.         [ Links ]

3. K. L. Gould., S.W. Casscells, L.M. Buja, D.C. Goff. The Lancet, 346, 750-753, (1995).         [ Links ]

4. J. Shepherd,. J. Cardiol. 82, 57T-59T, (1998).         [ Links ]

5. N.W. ShinichiroHiroshi, K. Kazuhiro, U. Shinya, H. Hideharu, O. Kyoichi, Hypertension Res., 28, 223-227, (2005).         [ Links ]

6. A. K. Robert and O. Albert J. Clin. Endocrinol. Metab., 87(2), 423-437,(2002).         [ Links ]

7. H Tajerzadeh, M Hamidi, J Pharm Biomed Anal 4,675-680, (2001)        [ Links ]

8. K. H. Yoon, W. Kim, J. Park and H. J. Kim, Bulletin of the Korean Chemical Society 6: 878-880, (2004).         [ Links ]

9. V.V. Pisarev1, N.E. Moskaleva1, Y.B. Zverkov1, L.B. Smirnova1, V.G. Belolipetskaya and Y.V. Sukhanov. Pharmaceutical Chemistry Journal, 2, 104-107, (2005).         [ Links ]

10 .M.P. Khalid, M. Muzeeb, S.S. Jafar, S.D. Basha, M. Ramesh and N. R. Srinivas. "Analysis of five HMG-CoA reductase inhibitors - atorvastatin, lovastatin, pravastatin, rosuvastatin and simvastatin: pharmacological, pharmacokinetic and analytical overview and development of a new method for use in pharmaceutical formulations analysis and in vitro metabolism studies". Biomedical Chromatography, 20 (3), 282 - 293, (2005).         [ Links ]

11. V. Shivva R. S Nikhil., R. K. Menon, C. A. Thammera, M. Ramesh and N.R. Srinivas "Simultaneous quantitation of rosuvastatin and gemfibrozil inhuman plasma by high-performance liquid chromatography and its application to a pharmacokinetic study". Biomed. Chromatography 20(11):1252-9, (2006).         [ Links ]

12. Thammera R. K., Nikhil R. S., Pasikanti K. K., Menon C. A., Venkata P. K., Ramesh M. and Srinivas N. R., "Determination of rosuvastatin in rat plasma by HPLC: validation and its application to pharmacokinetic studies". Biomed. Chromatogr.,20(9),881-887,( 2006).         [ Links ]

13. N. Sultana N, M.S. Arayne, S.Sajid S and S. Ali. Simultaneous Determination of Olmesartan Medoxamil and Irbesartan and Hydrochlorothiazide in Pharmaceutical Formulations and Human Serum using High Performance Liquid Chromatography Chinese Journal of Chromatography 26(5), 554-559, (2008).         [ Links ]

14. N. Sultana N, M.S. Arayne and B. Iftikhar Simultaneous determination of atenolol, rosuvastatin, spironolactone, glibenclamide and naproxen sodium in pharmaceutical formulations and human plasma by RP-HPLC Journal of the Chinese Chemical Society.55 (5), 1022-1029, (2008).         [ Links ]

15. N. Sultana N, M.S. Arayne, S. Naveed and H. Shamshad. An RP-HPLC method for simultaneous analysis of, and interaction studies on, enalapril maleate and H2-receptor antagonists. Acta Chromatographica 21 (4), 547-558, (2009).         [ Links ]

16. N. Sultana N, M.S. Arayne, N. Shafi N, F.A.Siddiqui and A. Hussain A Development of a RPHPLC method for the simultaneous analysis of diltiazem and statin: Application in pharmaceuticals and human serum Anal. Methods, 2, 1571-1576, (2010).         [ Links ]

17. N. Sultana N, M.S. Arayne and S. Naveed. Validated method for the simultaneous determination of lisinopril, pravastatin, atorvastatin and rosuvastatin in API, formulations and human serum by RP-HPLC Chinese Journal of Chemistry 29, 2421—2427, (2011).         [ Links ]

18. N. Sultana N, M.S. Arayne, S. Naz, N. Shafi and S. Naveed. Simultaneous determination of prazosine, atorvastatin, rosuvastatin and simvastatin in API dosage formulations and human serum by RP-HPLC. J Chinese Chemical Society 57(6), 1286-92, (2010).         [ Links ]

19. N. Sultana N, M.S. Arayne and W. Shehzad Simultaneous Determination of Ceftrioxazone sodium and Statin Drugs in Pharmaceutical Forms and Human Serum by RP-HPLC. Journal of The Chilean Chemical Society 55,(2), 193-198, (2010).         [ Links ]

20. N. Sultana N, M.S. Arayne, A.Z. Merza and H. Shamshad. High-performance liquid chromatographic analysis of pioglitazone, gliquidone, rosuvastatin and simvastatin in formulations and human serum Chin. J. Chem, 28(10), 1998-2002, (2010).         [ Links ]

21. ICH guideline Q2B; Validation of Analytical Procedures; Methodology(2003).         [ Links ]


(Received: March 18, 2010 - Accepted: May 30, 2011).